With three ranges of voltage with an accuracy range of 1 to 5 volts, you can check, calibrate and measure all your voltage, millivolt, and pH signal instruments. Save time calibrating; while other brands of calibrators require that you break the loop to calibrate voltage inputs, the PIE 235 simply clips the test leads across the voltage inputs in a live 4 to 20 mA loop without disconnecting any wires.
With three ranges of voltage with an accuracy range of 1 to 5 volts, you can check, calibrate and measure all your voltage, millivolt, and pH signal instruments. Save time calibrating; while other brands of calibrators require that you break the loop to calibrate voltage inputs, the PIE 235 simply clips the test leads across the voltage inputs in a live 4 to 20 mA loop without disconnecting any wires.
Features
General | |||
Operating Temperature Range | -20 to 60 °C (-5 to 140 °F) | ||
Common Mode Rejection | 50/60 Hz, 100 dB | ||
Size | 5.63 x 3.00 x 1.60 in, 143 x 76 x 41mm (L x W x H) | ||
Weight | 12.1 ounces, 0.34 kg (including boot & batteries) | ||
Batteries | Four "AA" Alkaline 1.5V (LR6) | ||
Battery Life | ≤ 50 hours | ||
Voltage Source | |||
Ranges and Resolution | -500.00 to 999.00 mV | ||
Accuracy | ≤ ± (0.02 % of Reading + 0.01% of Full Scale) | ||
Source Current | ≥ 20 mA | ||
Sink Current | > 16 mA | ||
Output Impedance | ≤ 0.3 Ohm at ≤ 20 mA load | ||
pH Source | |||
Accuracy in mV | ≤ %plusmn; (0.02 % of Reading in mV + 0.1 mV) | ||
Accuracy in pH | ≤ ± 0.003 pH @ 25°C |
Click on a category to view a selection of compatible accessories with the PIECAL 235 Process Voltage Calibrator, 20.000V Range.
General | |||
Operating Temperature Range | -20 to 60 °C (-5 to 140 °F) | ||
Common Mode Rejection | 50/60 Hz, 100 dB | ||
Size | 5.63 x 3.00 x 1.60 in, 143 x 76 x 41mm (L x W x H) | ||
Weight | 12.1 ounces, 0.34 kg (including boot & batteries) | ||
Batteries | Four "AA" Alkaline 1.5V (LR6) | ||
Battery Life | ≤ 50 hours | ||
Voltage Source | |||
Ranges and Resolution | -500.00 to 999.00 mV | ||
Accuracy | ≤ ± (0.02 % of Reading + 0.01% of Full Scale) | ||
Source Current | ≥ 20 mA | ||
Sink Current | > 16 mA | ||
Output Impedance | ≤ 0.3 Ohm at ≤ 20 mA load | ||
pH Source | |||
Accuracy in mV | ≤ %plusmn; (0.02 % of Reading in mV + 0.1 mV) | ||
Accuracy in pH | ≤ ± 0.003 pH @ 25°C |