Provides direct temperature calibration to all types of instruments such as transmitters, recorders, controllers, alarms, data acquisition, and computer systems. Also reads RTD & thermocouple outputs and displays temperature.
Provides direct temperature calibration to all types of instruments such as transmitters, recorders, controllers, alarms, data acquisition, and computer systems. Also reads RTD & thermocouple outputs and displays temperature.
Features
Operating Temperature Range | -20 to 60 °C (-5 to 140 °F) |
Storage Temperature Range | -30 to 60 °C (-22 to 140 °F) |
Temperature Effect | ≤ ±25 ppm/°C of Range Cold Junction Sensor: ≤ ±0.01% °/° from ambient temperature (°C or °F) |
Relative Humidity Range | 10 % ≤RH ≤90 % (0 to 35°C), Non-condensing 10 % ≤RH≤ 70 % (35 to 60°C), Non-condensing |
Normal Mode Rejection | 50/60 Hz, 50 dB |
Common Mode Rejection | 50/60 Hz, 120 dB |
Batteries | Four AA Alkaline 1.5V (LR6) |
Battery Life | 50 Hours |
Low Battery | Low battery indication with nominal 1 hour of operation left |
Protection against Misconnection | Over-voltage protection to 60 vrms (rated for 30 seconds) Red LED indicates OVERLOAD or out of range conditions |
Display | High contrast graphic liquid crystal display. LED backlighting for use in low lit areas |
Dimensions | 5.63 x 3.00 x 1.60" (143 x 76 x 41mm) |
Weight | 12.1 oz (0.34 kg) with boot & batteries |
Click on a category to view a selection of compatible accessories with the PIECAL 525B Automated Thermocouple & RTD Calibrator.
Operating Temperature Range | -20 to 60 °C (-5 to 140 °F) |
Storage Temperature Range | -30 to 60 °C (-22 to 140 °F) |
Temperature Effect | ≤ ±25 ppm/°C of Range Cold Junction Sensor: ≤ ±0.01% °/° from ambient temperature (°C or °F) |
Relative Humidity Range | 10 % ≤RH ≤90 % (0 to 35°C), Non-condensing 10 % ≤RH≤ 70 % (35 to 60°C), Non-condensing |
Normal Mode Rejection | 50/60 Hz, 50 dB |
Common Mode Rejection | 50/60 Hz, 120 dB |
Batteries | Four AA Alkaline 1.5V (LR6) |
Battery Life | 50 Hours |
Low Battery | Low battery indication with nominal 1 hour of operation left |
Protection against Misconnection | Over-voltage protection to 60 vrms (rated for 30 seconds) Red LED indicates OVERLOAD or out of range conditions |
Display | High contrast graphic liquid crystal display. LED backlighting for use in low lit areas |
Dimensions | 5.63 x 3.00 x 1.60" (143 x 76 x 41mm) |
Weight | 12.1 oz (0.34 kg) with boot & batteries |
Click on a category to view a selection of compatible accessories with the PIECAL 525B Automated Thermocouple & RTD Calibrator.