Three-in-one calibrator, calibrates thermocouple, RTD instruments and milliamp loops! Works like a 12 type RTD calibrator, 14 type thermocouple calibrator and can source 0 to 24 milliamps or simulate a two wire transmitter like a milliamp calibrator.
Three-in-one calibrator, calibrates thermocouple, RTD instruments and milliamp loops! Works like a 12 type RTD calibrator, 14 type thermocouple calibrator and can source 0 to 24 milliamps or simulate a two wire transmitter like a milliamp calibrator.
Calibrate all your Thermocouple, RTD instruments AND your milliamp loops with the PIE 525Plus. It has all the functions of a 12 type RTD calibrator, 14 type thermocouple calibrator and can source 0 to 24 milliamps or simulate a two wire transmitter like a milliamp calibrator. You may also perform stand alone calibrations of two wire transmitters by simultaneously supplying the RTD or thermocouple input signal while supplying the 24 volts to power up the transmitter and monitoring the milliamp output signal. Powers HART™ transmitters with built-in 250 ohm resistor simplifying hookups with HART communicators.
Features
Read mA | |
Ranges and Resolution | 0.000 to 24.000 mA or -25.00 to 125.00% of 4-20 mA |
Accuracy | ≤ ± (0.02 % of Reading + 0.003 mA) |
Source mA / Power & Measure Two Wire Transmitters & PWRM LEAK | |
Ranges and Resolution | 0.000 to 24.000 mA or -25.00 to 125.00% of 4-20 mA |
Accuracy | ≤ ± (0.02 % of Reading + 0.003 mA) |
General Specifications | |
Source Thermocouple Accuracy | ±(0.008% of Reading + 0.006 mV) |
Read Thermocouple Accuracy | ±(0.008% of Reading + 0.006 mV) |
Weight | 12.1 ounces, 0.34kg (including boot & batteries) |
Batteries | Four "AA" Alkaline 1.5V (LR6) |
Read mA | |
Ranges and Resolution | 0.000 to 24.000 mA or -25.00 to 125.00% of 4-20 mA |
Accuracy | ≤ ± (0.02 % of Reading + 0.003 mA) |
Source mA / Power & Measure Two Wire Transmitters & PWRM LEAK | |
Ranges and Resolution | 0.000 to 24.000 mA or -25.00 to 125.00% of 4-20 mA |
Accuracy | ≤ ± (0.02 % of Reading + 0.003 mA) |
General Specifications | |
Source Thermocouple Accuracy | ±(0.008% of Reading + 0.006 mV) |
Read Thermocouple Accuracy | ±(0.008% of Reading + 0.006 mV) |
Weight | 12.1 ounces, 0.34kg (including boot & batteries) |